| Color Display |
| Type |
Active matrix color LCD (TFT) with 17" viewing area |
| Resolution |
1280 x 1024 |
| Touch |
Full-screen, resistive |
| Functions |
User interface with touch-based icons and menus; Real-time waveform monitoring; Review previous waveform records while recording; Overlay numeric values in Engineering Units |
| Interface |
| Ethernet |
10/100BaseT |
| DVD |
Supports archive of data to
CD-R, CD-RW, DVD-R, DVD+R & DVD+RW |
| VGA |
For displaying data on an external monitor |
| USB 1.1 |
For external peripherals and file export |
| Analog Input Boards |
| Maximum Boards |
3 |
| Maximum Waveforms |
18 |
| A/D |
14 bit |
| Standard Event Inputs |
| Number of Inputs |
8 |
| Connector |
D-shell (9-pin) |
| Input Type |
TTL with pull-up (0 to 5VDC) |
| Response |
Detects if duration is greater than 0.01 msec |
| Real-time Signal Processing |
| Filters |
Low Pass Stops: 10 Hz to 10,000 Hz; High Pass Starts: 0.1 Hz to 100 Hz; Notch: 50 or 60 Hz center |
| RMS |
Time constant selectable from 0.02 to 2 seconds |
| Frequency to Voltage |
5 Hz to 20 kHz, +/- 0.05% of measurement + 0.1 Hz |
| Cross-Channel |
Sum, Difference, Ratio, Product (Power) |
| Real-time Signal Testing |
| Basic |
All active signals monitored simultaneously |
| Test Types |
Tests include window, slope/level, slew and event pattern |
| Outputs |
Result available for alarms, trigger, abort and external TTL |
| Data Capture |
| Recording Method |
Internal 73 Gbyte hard drive |
| Max. Rate |
100,000 samples/second per channel (all channels) |
| Min. Rate |
1 sample/minute |
| Dual Rate |
Second slower rate available when no trigger is used |
| Total Capacity |
16 billion samples |
| Max. Record |
2 billion samples |
| Header |
Information on units, range, sample rates, etc. saved with data |
| Events |
All standard event inputs captured with waveforms |
| Trigger Point Lock |
Amounts of pre- and post-trigger is user adjustable |
| Auto Re-arm |
Allows automatic stacking of captures |
| Auto Playback |
Yes |
| External Sample Rate |
TTL sample clock up to 40,000 Hz |
| Signal Conditioner Specifications (General) |
| Channels per Module |
6 |
| Isolation |
250 Vrms (iso-common to chassis and other iso-commons) |
| Bandwidth |
12 kHz (-3dB) |
| Input Coupling |
DC |
| Zero Suppression |
Yes |
| User Engineering Units |
Yes |
| Calibration |
Semi-automated to external reference |
| Single Ended Voltage Input |
| Input Type |
Isolated, single-ended |
| Connector |
Guarded banana jack |
| Max. rated input |
+/- 250 Vrms |
| Max. transient input |
+/- 400 V |
| Specified ranges |
40 to 400 VFS; 4 to 40 VFS; 0.4 to 4VFS |
| Accuracy (25°C) |
+/- 0.5% of attenuator |
| Min. Input Impedance |
1 Megohm |
| Differential Voltage Measurements (use for bridge measurements) |
| Input Type |
Isolated, differential |
| Connector |
Screw terminal header |
| Absolute Max. Input |
+/- 40 V differential |
| Measuring Ranges |
200 to 1600 mVFS; 50 to 500 mVFS; 5 to 50 mVFS |
| Accuracy (25°C) |
+/- 0.5% of attenuator |
| Min. Input Impedance |
200 kOhm |
| CMR at 60 Hz |
> 60dB |
| Auto Balance |
Yes, limited to maximum span |
| Thermocouple Measurements |
| Input Type |
Isolated, differential with screw terminal header connector |
| Absolute Max. Input |
+/- 40 V |
| Specified Range |
Type J: 0 to 760 °C
Type K: 0 to 1370 °C
Type T: -160 to 400 °C
Type E: -100 to 870 °C
Type N: 0 to 1300 °C |
| Accuracy (25°C) |
0.5% of measurement + 2 °C |
| Bandwidth |
10 Hz (-3dB) |
| Intrinsic Noise |
0.5 °C (1 Hz filter) |
| Linearization |
NIST polynomial |
| Units |
°C or °F |
| Excitation |
| Connector |
Screw terminal header |
| Excitation |
Isolated 10V @ 20 mA |
| Adjustable |
0.1 to 10 V |
| Frequency Measurements |
| Method |
Zero-crossing, channels 1, 7 and 13 only |
| Signal Amplitude |
Peak-to-peak minimum is 20% of attenuator |
| Hysteresis |
Yes |
| Removable Drive |
| Type |
DVD+R |
| Format |
MSDOS® format |
| Function |
Setup files, software upgrades, data transfer/archive |
| Power |
| Input voltage range |
102 to 264VAC |
| Frequency Range |
47 Hz to 63 Hz |
| Power Consumption |
300 W maximum |
| Compliance |
| Safety |
EN 61010-1 |
| EMC |
FCC Class A, EN 61326 |
| Power Harmonics |
IEC1000-3-2 |
| Physical |
| Enclosure |
Aluminum |
| Dimensions |
16.1" L x 15.1" W x 6.4" H |
| Weight |
29 lbs. |
| Environmental |
| Operating Temperature |
5 to 40 °C (40 to 105°F) |
| Non-operating Temperature |
-20 to 60 °C (-4 to 140 °F) |
| Operating humidity |
10% to 90%, non condensing |
| Optional Chart Recorder - Model SCR-18 |
| Chart Width |
11" |
| Max. Chart Speed |
50 mm/sec |
| Optional DC Input Power - Model PIM-18 |
| Input Range |
9-18 VDC |
| Input Power |
200 W max |
| Fuse |
User replaceable 25 A |
| Optional DC Input Power - Model PIM-18A |
| Input Range |
20-32 VDC |
| Input Power |
240 W max |
| Fuse |
User replaceable 10 A |